Human IL-12B p40 ELISA Kit
pg/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0228 |
0.0235 |
0.0232 |
|
1.37 |
0.0310 |
0.0304 |
0.0307 |
0.0076 |
4.12 |
0.0481 |
0.0481 |
0.0481 |
0.0250 |
12.35 |
0.0945 |
0.0931 |
0.0938 |
0.0707 |
37.04 |
0.2360 |
0.2307 |
0.2334 |
0.2102 |
111.11 |
0.6390 |
0.6322 |
0.6356 |
0.6125 |
333.33 |
1.7120 |
1.6570 |
1.6845 |
1.6614 |
1000.00 |
3.8910 |
3.8380 |
3.8645 |
3.8414 |
Intra-assay Precision |
Inter-assay Precision |
|||||
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
22 |
22 |
22 |
6 |
6 |
6 |
|
Average(pg/ml) |
19.4 |
104.0 |
339.2 |
19.3 |
101.1 |
309.6 |
Standard Deviation |
0.7 |
4.3 |
14.4 |
1.3 |
5.4 |
20.1 |
Coefficient of Variation(%) |
3.4 |
4.1 |
4.2 |
6.6 |
5.4 |
6.5 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The recovery ranged from 90% to 117% with an overall mean recovery of 102%.
Sample Matrix |
Sample Evaluated |
Range (pg/ml) |
Detectable (%) |
Mean of Detectable (pg/ml) |
Serum |
30 |
25.15-193.73 |
100 |
80.27 |
Serum/Plasma – Thirty samples from apparently healthy volunteers were evaluated in this assay. Donor medical histories were not available.