Mouse PD-1 ELISA kit
pg/ml |
O.D. |
Average |
Corrected |
|
0.00 |
0.0264 |
0.0261 |
0.0263 |
|
13.72 |
0.0304 |
0.0335 |
0.0320 |
0.0057 |
41.15 |
0.0453 |
0.0421 |
0.0437 |
0.0175 |
123.46 |
0.0771 |
0.0764 |
0.0768 |
0.0505 |
370.37 |
0.1637 |
0.1724 |
0.1681 |
0.1418 |
1111.11 |
0.4431 |
0.4195 |
0.4313 |
0.4051 |
3333.33 |
1.2060 |
1.2660 |
1.2360 |
1.2098 |
10000.00 |
3.2720 |
3.0420 |
3.1570 |
3.1308 |
Intra-assay Precision |
Inter-assay Precision |
|||||
Sample Number |
S1 |
S2 |
S3 |
S1 |
S2 |
S3 |
22 |
22 |
22 |
6 |
6 |
6 |
|
Average(pg/ml) |
187.2 |
963.7 |
3386.5 |
201.2 |
1007.7 |
3360.1 |
Standard Deviation |
10.5 |
69.0 |
211.2 |
4.4 |
35.5 |
223.5 |
Coefficient of Variation(%) |
5.6 |
7.2 |
6.2 |
2.2 |
3.5 |
6.7 |
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.
The recovery ranged from 94% to 84% with an overall mean recovery of 87%.
Sample Matrix |
Sample Evaluated |
Range (pg/ml) |
Detectable (%) |
Mean of Detectable (pg/ml) |
Serum |
30 |
133.17-389.78 |
100 |
261.5 |
Serum/Plasma – Thirty samples from apparently healthy mice were evaluated in this assay. Donor medical histories were not available.