Mouse PD-1 ELISA kit

Key features and details

  • Specification: 96 Test
  • Sensitivity: 7.65 pg/ml (10 μl)
  • Standard Curve Range: 13.72~10000 pg/ml
  • Standard Curve Gradient: 7 Points/3 Folds
  • Number of Incubations: 2
  • Sample Volume: 50 μl/10 μl
  • Assay type: Sandwich Elisa
  • Operation Duration: 120min
  • Brand:
CAT.NO. : AEM0037
RMB Please choose
RMB Please choose
Size:
96T
Trail, Bulk size or Custom requests Please contact us
Product Details
Background
PD-1 (Programmed Death-1 receptor) is a key regulator of the threshold of immune response and peripheral immune tolerance. It is expressed on activated T cells, B cells, monocytes, and dendritic cells and binds to PD-L1 or PD-L2. PD-1 ligation induces co-inhibitory signals in T cells promoting their apoptosis, anergy, and functional exhaustion. Blockade of the PD-1: PD-L1 interaction using anti-PD-1 antibodies enhances understanding of anti-tumor immunity and advances the potential for cancer immunotherapy.
Typical Data


pg/ml

O.D.

Average

Corrected

0.00

0.0264

0.0261

0.0263

13.72

0.0304

0.0335

0.0320

0.0057

41.15

0.0453

0.0421

0.0437

0.0175

123.46

0.0771

0.0764

0.0768

0.0505

370.37

0.1637

0.1724

0.1681

0.1418

1111.11

0.4431

0.4195

0.4313

0.4051

3333.33

1.2060

1.2660

1.2360

1.2098

10000.00

3.2720

3.0420

3.1570

3.1308

Precision

Intra-assay Precision

Inter-assay Precision

Sample Number

S1

S2

S3

S1

S2

S3

22

22

22

6

6

6

Average(pg/ml)

187.2

963.7

3386.5

201.2

1007.7

3360.1

Standard Deviation

10.5

69.0

211.2

4.4

35.5

223.5

Coefficient of Variation(%)

5.6

7.2

6.2

2.2

3.5

6.7

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays) Three samples of known concentration were tested six times on one plate to assess intra-assay precision.

Spike Recovery
The spike recovery was evaluated by spiking 3 levels of mouse PD-1 into health mouse serum sample. The un-spiked serum was used as blank in this experiment.
The recovery ranged from 94% to 84% with an overall mean recovery of 87%.
Sample Values

Sample Matrix

Sample Evaluated

Range (pg/ml)

Detectable (%)

Mean of Detectable (pg/ml)

Serum

30

133.17-389.78

100

261.5


Serum/Plasma – Thirty samples from apparently healthy mice were evaluated in this assay. Donor medical histories were not available.
New Products
Get in touch with AREX
Name:*
Tel/Phone:*
Company:*
Email:*
Inquiry:
Captcha*
Submitting your email information means that you are willing to receive email information from AREX regarding technology, applications, products, and events. By clicking on the 'unsubscribe' button in the email or by contacting info@arexbio.com You can unsubscribe at any time by sending an email. Regarding your data usage information, please refer to our privacy policy.